Amplifier AP5200-8

Purpose:
  • Vibration (vibroacceleration) measurements in the modes:
  • charge amplifier;
  • voltage amplifier.
Features:
  • Vibroacceleration measurements.
  • Normalization mode.
  • Overload indication.
  • Low level of noise.
  • Built-in voltmeter that measures:
  • RMS of vibroacceleration in charge amplifier and voltage amplifier modes.
  • Touch-sensitive keyboard.
  • Casing 19": metal, height: 3U, depth: 260mm.
Check price with the manufacturer
Description
Documents
  • Parameter
    Value
  • Maximum input charge (peak value)
    ± 100000 pC
  • Maximum input voltage (peak value)
    ± 10 V
  • Conversion factor for charge (corresponds to a series of 1, 2, 5)
  • Conversion factor for voltage (corresponds to a series of 1, 2, 5)
  • Normalization range (increment 0.01) of conversion factor
  • Limits of basic relative error of conversion factor setting at 1 kHz
    ± 0,8 %
  • ± 1,6 %
  • Maximum output voltage (peak value) at nonlinear distortion factor < 5%, not lower than
    ± 10 V
  • Operating frequency range with attenuation on the boundaries -10 %
    from 0,3 to 100000 Hz
  • from 1 to 2000 Hz
  • Input resistance (higher than)
    10000000 кОm
  • Output resistance (not higher than)
    100 Оm
  • Acceptable additional relative error limits of conversion factor for charge in the operating temperature range 0 to 50 °C
  • Cutoff frequencies of built-in high pass filters (increment 0.1; 1; 10; 100 Hz) with decline in frequency response not less than 80dB/decade and attenuation -10 %
    from 0,3 to 200 Hz
  • Cutoff frequencies of built-in low pass filters (increment 0.1; 1; 10; 10 kHz) with decline in frequency response not less than 80dB/decade and attenuation -10 %
    from 0,1 to 100 кHz
  • Basic relative error limits for RMS value of the output signal voltage in the range from 100 mV to the maximum and within frequencies from 20 to 100 000 Hz
    ± 1 %
  • + (12 ± 2) V
  • Powering mode of sensors
    + (24 ± 2) V
  • 5,7 ± 0,7 V
  • Root mean square (RMS) value of noise reduced to the input, in the mode of voltage conversion (not higher than)
    30 µV
  • Root mean square (RMS) value of noise reduced to the input, in the mode of charge conversion for capacitance 1 nF (not higher than)
    0,03 pC
  • Run mode setting time after power-on or switching inputs and conversion factors (not longer than)
    5 s
  • Power supply
    mains supply from adapter
  • Connector type
    input / output
  • Overall dimensions
    157 x 232 x 75 mm
  • Weight
    2000 g
  • Delivery set
    The amplifier of the measuring AP5200; Passport; operator's Manual; AR10 Cable adapter; Cable adapter AR15; Setup CD-ROM with the program AP5110 Explorer; USB Cable A-B; user Manual; power Unit AS05-01 (~220 V/=12 V, >= 1 ) ; Harness RS-485